FEATURES
Flow Rate 0.1 To 1200 Ml/Hr
Purge Up To 1200 Ml/Hr
Bolus Rate Up To 1200 Ml/Hr
Pressure Limits Settable At Low / Mid / High
Alarms: Expressed By Means Of Flashing Message On The Display And Sound Beep
Set Installation: Door Open, Occlusivity Check
Device Control: No Mains, Battery Pre-Alarm & Alarm, Idle,Technical Malfunction, Unconfirmed Programming, Preventive Maintenance
Infusion Control: Occlusion Upstream & Downstream, Infusion End Pre-Alarm & Alarm, Kvo/Kor, Container Empty
Keypad Locking: Restricts Accidental Change Of The Infusion Parameters
PRE ALARMS/ ALARMS/ SECURITY FEATURES
Alarms : Expressed by means of flashing message on the display and sound beep
Device Control : No Mains, Battery Pre-alarm & Alarm, Idle, Technical Malfunction, Unconfirmed Programming Preventive Maintenance
Infusion control : Occlusion Upstream & Downstream, Infusion End Pre-alarm & Alarm, KVO/KOR, Container Empty
Preventive Control : Door Open Alarm with door locking feature Air in Line (Ultrasonic Air Bubble Detector)
Keypad Locking : Restricts accidental change of the infusion parameters
STANDARD ACCESSORIES
Unit with pole clamp, power cord & operating manual
OPTIONAL / UPGRADABLE FEATURE
Power Source : External power 12V DC Converter (For Ambulatory Power Supply)
Battery Life : Upgradable to 5 / 10 hr @ 25 ml/hr
Stackability : Multiple input AC Cord To Power Up To 8 Pumps
Communication Port : Rs232 Port
Data Log Event : 500 Data Log Events In Real Time Expandable Up To 1500 nos.
Other : Nurse Call Attachment
BASIC INFUSION CONTROLS
IV Line Brands : Bbraun, Jms, Romson, Polymed (Provision To Calibrate 13 More Brands
Flow Rate : 0.1 To 1200 Ml/Hr (Increment In Step Of 0.1 Ml/Hr) Facility For Online Titration
Purge : Up To 1200 Ml/H
Priming : 3 Modes: Mandatory/ Non Mandatory/ Advised Rate : Up To 1200 Ml/Hr
Bolus Type : Auto & Manual
Bolus Rate : Up To 1200 Ml/Hr
Bolus Volume : 0.1ml To Max Capacity Of Syringe (0.1ml Incr.
Bolus Indicator : Beep After Infusion Of Every 0.5/1 Ml Bolus Volume
Infusion Modes * : 1] Rate Ml/Hr (0.1ml/Hr Incr.) 2] Volume Limit 0.1-999.9 Ml (01 Ml Incr.) 3] Volume/Time 0.1-999.9ml; 00h01-100h00 4] Dose Rate Dose Rate # Dose Unit :Mcg/Min, Mcg/Hr, Mcg/Kg/Min Mcg/Kg/Hr Mg/Min, Mg/Hr Mg/Kr/Min, Mg/Kg/Hr #Dilution :X Mg/Ml Or X Mg / Y Ml
Stand by : Programmable Form 1 Min. To 24 Hrs
Drug Labeling : 64 Drug Name With Provision For User Add/Delete
Drug Library : In Built ‘prudent’ Drug Library Editor For Setting Normal Infusion, Bolus And Loading Dose Parameters. Adjustment Of The Dilution, Default Infusion Modes, Flow Rate & Bolus Limits In Ml/H
KVO/KOR * : Settable From 0.1 To 5 Ml/Hr Provision To Keep It Off.
Overall Accuracy : +/- 5% (Measured Time.1hr, Min 0.2 Ml) (Measured As Per Iec 60601-2-24)
Plenumtek Infusion Pump
Easy Return And Exchange Policy
Non Returnable
Installation
No Required
Product Condition
New
Brand
Plenumtek
Warranty Policy :
One Year
Delivery Time :
8 To 12 Day's
TECHNICAL SPECIFICATION
WEIGHT
500 g
BRAND
Plenumtek
DEPARTMENT
ICU & Critical Care
MASTER CATEGORY
Equipment
PRODUCT CATEGORY
Infusion Pump
PRODUCT TYPE
Volumetric Infusion Pump
PRODUCT SPECIFICATION
Product Type
Volumetric Infusion Pump
Pumping Mechanism
Linear Peristaltic
Dimension
130mm (W) x 183mm (L) x 115mm (H)
Weight
2.2 Kgs
Display Size & Type
66mm x 33mmm Graphic LCD Monochrome
Parameters Displayed
Occlusion Level, IV Set Brand, Drug Name, Set Flow Rate, Infused Volume, Battery Life, Alarms & Indicators
Keypad
Capacitive Touch Screen Keypad
Fixation
Allows Horizontal & Vertical Mounting
Stackability
For interlocking of two or more units
Power Source
110/220 VCA (+/- 20%), 50/60 Hz, 12 VA
Battery
7.2 Rechargeable NiMh Capacity minimum 3.5 hrs @ 25 ml/hr * Recharging time 16 hrs
Battery Status
Display in hrs. and min.
Power Saver
Auto adjust display brightness to save battery
Operating Temp.
05 to 40 C
Relative Humidity
15 of 90% (no condensation)
Atmospheric Pressure
70 to 106 Kpa
PRESSURE MANAGEMENT
Pressure Modes
3 Pre - set Levels
Pressure Limits
Low 0.4 kg/cm (300 mmHg)
Mid 0.8 kg/cm2 (600 mmHg)
High 1.2 kg/cm (900 mmHg)
Tolerance
+/- 20%
DPS
Warns pressure variation