DESCRIPTION
The Philips Goldway Gs20 Is A Top-Of-The-Line Patient Monitor That Provides Accurate Details And Actionable Patient Data During Patient Monitoring With Its 10-Inch Display, The Philips Gs20 Patient Monitor Is Easy To Use And Provides Clinicians With All The Information They Need To Make Informed Decisions About Their Patients' Care. The Philips Gs20 Patient Monitor Offers Essential Measurements For Daily Use, Such As The Oxycrg And Mini-Trend Software Functions That Provide Basic Clinical Support It Can Be Connected To The Philips Efficia Cms200 Central Station And Includes An Hl7 Data Output Function For Emr/His Integration Moreover, The Monitor Features The Philips Fast Sp02 And Philips ECG Algorithms
FEATURES
ADVANCED FUNCTIONALITIES
Philips Gs20 Patient Monitor Provides Real-Time Data On Cardiac And Respiratory Conditions In Order To Fit Different Clinical Settings As Well As Specific Monitoring Parameters Such As Philips St.) Ar 3-And5-Lead Ecg/ Respiration, Nbp 2 Temperatures And Philips Spo2/Pr St And Arrhythmia Monitoring, It Also Offers Options For 2-Channel Intrusive Blood Pressure And Microstream Etc02 Monitoring
EASE OF USE
Gs20 Patient Monitor Has A 10.4-Inch Color Display With Flexible Display Modes Such As Large Number Display Minitrend, Oxycrg, And The Neight Adjustment Function Of The Auto Waveform Channel To Fit Different Clinical Settings For Maximum Ease Of Use Intuitive Menus, Fixed Key Access Bright Color Screen, And Optional Touch Screen Nav All Been Designed
COMPACT AND COST-EFFECTIVE
The Philips Goldway Gs20 Patient Monitor, Designed In Compliance With Philips Exacting Quality Standards, Offers Market-Proven Critical Parameters In A Lightweight Cost-Effective Display To Produce Precise Measurement Results On Your Tight Budget.
WORK WITH EFFICIA CONSUMABLES
The Philips Gs20 Patient Monitor Is Designed To Be Compatible With Consumables From Philips Efficia, Which Are Also Used On Patient Monitors From The Philips Emicis Cm Series, To Reduce Your Total Cost Of Ownership By Sharing Consumables From Philips Across Various Care Settings
SPO2
- Measurement Range : 0- 100%
- Accuracy : 70 % - 100 % , Depends On Sensor , 0 -69 % , Unspecified
- Waveform Speed : 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
- Pulse Rate Range : 30 bpm -300 bpm
- Pulse Rate Accuracy : 土2 bpm or 土2%, Whichever is Greater
- Perfusion Index : Yes
RESPIRATION (RESP)
- Measurement Method : Thoracic Impedance
- Measurement Range : 0 rpm - 150 rpm
- Accuracy : 0 rpm - 120 rpm : 土 1 rpm ; 121 rpm - 150 rpm: 土2 rpm
- Waveform Speed : 6.25 mm/s, 12.5 mm/s, 25 mm/s , 50 mm/s Accuracy Undefined
- Gain Selection : x ½ , x 1, x 2
- Apnea Alarm : Yes , 10-60 s
- Work Mode : Auto Or Manual (Threshold)
- RESP Lead : RA-LL
TEMPERATURE (TEMP, 2- CHANNEL)
Measurement Range : 0 -50 °C (32 °F - 122 °F)
Accuracy : 土0.1 °C (土0..2 °F ) ( Without Sensor)
Resolution : 0.1 °C (0.2°F)
Units : °C / °F
Minimum Measuring Time : Attachable skin Surface Probe Less Than 10 Minutes Esophageal/Rectal Probe : Less Than 2 Minutes b
Philips Goldway GS20 Monitor
Easy Return And Exchange Policy
Non Returnable
Installation
No Required
Product Condition
New
Brand
Phillips
Warranty Policy :
One Year
Delivery Time :
8 To 15 Day's
ECG
Lead Type
3-Lead Or 5-Lead Selectable
Lead Selection, 3-Lead
I, II, III
Lead Selection, 5-Lead
I, II, III, aVR aVL, aVF. V
Hr Measurement
Adult/Pediatric: 15 bpm-300 bpm Neonatal 15 bpm - 300 bpm Step 1 bpm
Hr Accuracy
15 bpm or ±1%, Whichever is
Arrhythmia
21 calls
St Measurement Range
-25 mV-+2 5 mV
Waveform Speed
6.25 mm/s 12.5 mm/s, 25 mm/s,
Gain Selection
x 1/4, x 1/2 x 1, x2, x 4, Auto
Pacemaker Detection
Yes
NON-INVASIVE BLOOD PRESSURE (NIBP)
Method
Automatic Oscillometic
Parameters
SYS/DIA/MAP/Pulse Rate
Operating Mode
Manual/Auto/STAT
Auto Measure Intervals
3, 5, 10, 15, 30, 60, 90, 120, 240 480 Minutes
Units
mmHg/kPa
Cuff Pressure Range , Adult/ Pediatric
0-300 mmHg (0 kPa -40.0 kPa)
Cuff Pressure Range , Neonatal
0-150 mmHg (0 kPa - 20.0 kPa)
Systolic Measurement Range, Adult /Pediatric
Systolic Measurement 30mm rig-135 mmHg (4 0 kPa -33.9 kPa)
Systolic Measurement Range, Neonatal
30 mmHg -135 mmHg (4.0kPa-18.0 kPa)
Diastolic Measurement Range , Adult / Pediatric
10 mmHg -220 mmHg (1.3 kPa-29.3 kPa)
Diastolic Measurement Range , Neonatal
10 mmHg -110 mmHg (1.3 kPa-14.7 kPa)
Map Measurement Range , Adult / Pediatric
20 mmHg -235 mmHg (2.7 kPa-31.3 kPa)
MAP Measurement Range ,Neonatal
20 mmHg -125 mmHg (2.7 kPa-16.7 kPa)
Accuracy , Maximum standard Deviation
8 mmHg (1.1 kPa)
Accuracy , Mean Error
土5 mmHg (土0.7 kPa)
Pulse rate Range
40 bpm -240 bpm
Pulse Rate Accuracy
土 5 bpm or 土5%, Whichever is Greater