top of page

SHORT SPECIFICATION

 

  • ETO Cartridge: 100 Gms -1 Nos
  •  Fully Automatic
  •  PLO Operated
  •  Chamber SS 316 (3mm)
  •  Touch Screen HMI 7 Inch
  •  Easy And Cost Effective

 

ADVANTAGES OF PLC AUTOMATION                                                                                    A Programmable Logic Controller, PLC Or Programmable Controller Is A Digital Computer Used For Automation Of Electro- Mechanical Processes, Such As Control Of Machinery. 

1. Pre-Programmed Cycles: Sterilization Of Varied Equipment’s Like Gloves, Linen, Gauze, Cotton, Instruments, Glassware Etc. Preprogrammed In The PLC

2. Control: During Sterilization, PLC Controls The Temperature, Moisture, Valve Control, And All Critical Time And Pressure-Based Parameters. 

3. Temperature : Is Maintained Uniformly & Accurately (± 0.5°c). Continuous Display Of Temperature, Pressure, And Humidity On-Line Always Allows The User To Be Aware Of Cycle Status. 

4. Alarm: In Case Of Any Error, An Error Code & Alarm Is Generated, And The Cycle Stops Automatically. Computer Screen Displays Error Message And The Possible Cause & Remedy. 

5. Print: Users Can Reprint Reports As And When Required Using Any Printer. Digital Data Cuts Cost & Space, Ensures Safety Of  Data, And Negates The Cost Of Expensive Thermal Printers. 

6. HMI: The Operator Can Run A Machine Through HMI (Human Machine Interface). He Can Also Change The Programs According To The Hospitals Suitability. 

7. Security: Advanced Multi-Level Password Protection Enhances Security. This Avoids Unauthorized Access.

 

DESCRIPTION OF THE AUTOMATION

The Machine Is Provided With Microprocessor / PLC Based Automation Along With An LCD/ Touch-Screen Display, Which Allows The Operator To 

  •  Change And Set The Parameters        

  •  Run The Cycle According To Pre-Selected Cycles.

  •  Monitor Various Phases / Individual Parameters During The Process.

  •  View On/Off Status Of The Various Components Of The Machine While The Cycle Is In Progress. 

                                                                                                                                                

Software For Controlling, Monitoring, And Recording Of Various Parameters Is Provided. This Makes It Possible To Operate And Monitor The Machine By Connecting It With Any Computer Where The Software Is Loaded, Apart From Running It From The Machine Itself.  

The Above Devices Control And Regulate All The Functions Of The Machine. It Makes The System User Friendly And Allows The Operator To Change The Process Parameters As And When Necessary.  

 

THE ADVANTAGE OF THE ABOVE IS THAT THE SYSTEM CAN CARRY OUT MULTIPLE PROGRAMS

 

IT INCLUDES FEATURES LIKE 

  •  Password Protection For Unauthorized Access. 

  • Analog Input Of Process Pressure/Vacuum. 

  •  Digital/Analog Input For Temperature. Temperature Is Indicated In The Sterilization Area, By The Use Of A Temperature Sensor. 

  • One Relative Humidity And One Pressure/Vacuum Sensor For Display Of The Humidification Efficacy And Current Pressure/Vacuum, That The Load Is Subjected To, At All Times During The Process, Is Displayed In Graph Format When The Process Is Going On And Is Also Available In Report Form For Printing At The End Of The Cycle. 

  • Humidity Sensor Is Movable Inside The Chamber And Hence A Comprehensive Mapping Of All Areas Of The Chamber Is Possible.

  •  PC Backup System In Case Of HMI Failure And Vice-Versa.  

 

SAFETY 

  • The Cycle Operates In Vacuum And Hence Is Safer To Operate, As There Is Minimum Chance Of Leakage.

  • The Electronic Design Provides Accuracy Of Operation, Dependability And Repeatability.

  • The Sterilizer Can Aerate Items After Sterilization. The Sterilization-Aeration Process Can Be Continuous In One Chamber And Hereby Eliminate Gas Exposure During Load Transfers To Aerators.

  • After The Sterilization Cycle, The Chamber Is Purged Or Aerated Continuously Until The Door Is Opened. This Prevents A Buildup Of Ethylene Oxide From Items Aerating In The Chamber. 

 

OPTIONAL ACCESSORIES

  •  Air Compressor

  •  Hand Sealing Machine

  •  ETO Aluminum Cartridges (40 Gms)

  •  ETO Packaging Roll

ETO 12"x12"x48"( 4cft Capacity,110 Liter)

₹550,000.00 Regular Price
₹495,600.00Sale Price
Taxes Included |
  • Easy Return And Exchange Policy

    Non Returnable 

  • Installation

    Free

  • Product Condition

    New

  • Brand

    Honmed

  • Warranty Policy :

    One Year

  • Delivery Time :

    15 To 25 Day's 

  • TECHNICAL SPECIFICATION

     

                                      ITEM 

                                DESCRIPTION 

    Inner Seal (Chamber) 

    SS 316L Thickness: 5mm

    Chamber Operating Pressure 

    Negative Pressure Cycle

    Chamber Testing Pressure 

    1.5 Kg/Cm2 

    Single Door 

    SS 316L  Thickness: 10mm

    Gasket

    Silicon Gasket (10/12 MM Solid Section)

    Steam Generator 

    SS 316L Thickness: 3mm

    Outer Cover 

    Factory Finished (Poly-Coated) MS

    Heating 

    SS 316L Heaters On All Sides Of The Chamber For Uniform Heating Throughout The Load.

    Temp Sensor 

    For Accurate Measuring And Controlling Of Temp. In The Chamber

    Operating Temp 

    Maximum: 45°c To 55°c (Chamber)  

    (1) Medium Cycle: 35°c To 45°c 

    (2) Cool Cycle: 35°c To 40°c

     (3) Hot Cycle: 48°c To 55°c 

    RH % Sensor 

    For Measuring And Controlling The Humidity Of The Chamber

    Pressure Transmitter 

    For Measuring And Controlling The Pressure-Vacuum That The Load Is Subjected To During Removal And Charging Of Gas. 

    All Fittings On The Machine

    SS 304 

    All Valves For Automation

    Teflon Construction Pneumatic Valves, Operated By Comp. Air Supply Through Solenoid Valves.  

    Humidification 

    Special Humidifier For The Purpose Of Humidification Of The Load And Heating Of The Gas.  

    Vacuum 

    Vacuum Created By The Venturi System. (Make: Festo)

    Automation Hardware 

    PLC Make: Kinco 

    HMI Make: Kinco

    Process Reports With Thermal Printer 

    Reports Are Generated By The Software For Each And Every Cycle In Three Formats, Viz. (I) Single Page Summary Report And (Ii) Detailed Report 

    Testing 

    Pressure / Vacuum Testing As Per Specified Standards. (1.5 Times The Working Pressure) 

    Operation 

    All Functions Are Electro-Pneumatically Operated. Door Closing Is Manual

    Loading Trolley 

    Fabricated With SS 304

    Battery Backup

    Ups Systems Will Be Provided For Battery Backup. If Power Fails It Will Be Usable For 20 Minutes For Continuous Power Supply.  

    Air Filter 

    0.2 Micron Air Filter Is Also Provided With Machine For Fresh Air. 

bottom of page