top of page

SHORT SPECIFICATION 

 

  •  Drop Mode
  •  Time Mode
  •  Bolus
  •  1 to 2 Hours Battery

 

FEATURES 

 

  • HD LCD Display, High Capacity Words, Friendly User Interface, Dynamically Display Working Status;

  • Audible And Visual Alarm For Occlusion, Empty, Low Battery, End Of Infusion, Door Open, Wrong Setting, Etc., Which Gain Patents;

  • Compatible With Any Brands Of Infusion Sets After Correct Calibration;

  • Preset Solution Volume To Greatly Reduce The Workload Of Nurses;

  • Work Mode: Ml/H And Drop/Min Can Switch Freely;

  • Three Levels Of Occlusion: High, Middle And Low;

  • Purge Function;

  • Kvo (Keep-Vein-Open) Automatically Opens As Infusion Is Completed, Kvo Rate Is 1-5 Ml/H(1ml/H Step);

  • Power Source: Ac100---240v, 50/60hz; Internal Battery;

  • Automatically Record The Settings Of Last Infusion;

  • Specification Of Infusion Pump 

 

EASY TO CARRY

 

  • Support Ml/H And D/Min Mode And Can Change By One Key

  • German Import Silent Motor Ic And Japan Motor,Low Noise, Low Vibration                                     

  • Start With Self Checking Function Makes Stably System Running

Byond Infusion Pump

₹32,000.00 Regular Price
₹23,520.00Sale Price
Taxes Included |
  • Easy Return And Exchange Policy

    Non Returnable 

  • Installation

    No Required 

  • Product Condition

    New 

  • Brand

    Byond 

  • Warranty Policy :

    One Year

  • Delivery Time :

    8 To 12 Day's 

  • TECHNICAL SPECIFICATION

     

    Rate 

    1 ml/h ~ 1,200ml/h 

    Flow Rate Accuracy 

    Within ± 5% (after correct calibration) 

    Mechanical precision 

    ±2% 

    Purge Rate 

    100ml/h ~ 1,000ml/h (100 ml/h step) 

    Infusion Volume 

    1ml ~ 9999ml 

    Total Infusion Volume 

    0.1ml ~ 9999.9ml 

    KVO rate 

    1ml/h ~ 5 ml/h (1ml/h step) 

    Occlusion 

    High: 800mmHg ± 200mmHg (106.7kPa±26.7kPa 

    Medium: 500mmHg ±100mmHg(66.7kPa±13.3kPa 

    Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa ) 

    Unique Audible And Visible Alarm 

    Human Voice Alarm For End Of Injection, Occlusion, Door Open, Bubbles In The Tube, Wrong Setting, Low Battery, Ac Power Pulled Out Etc. 

    Power Source 

    Ac 100v ~ 240v, 50/60hz; Internal Rechargeable Li Battery, Capacity ;1,600mah, 4 Hours Internal Battery Backup 

    Bubble Detector 

    Ultrasonic Wave Detector; Detection Sensitivity≥ 25μl 

    Fuse 

    F1al/250v, 2 Pc Inside 

    Power Consumption 

    18va 

    Operating Condition 

    Ambient Temperature: +5℃ ~ +40℃; 

    Relative Humidity: 20 ~ 90% 

    Atmospheric Pressure: 86.0kpa ~ 106.0kpa 

    Transport & Storage Condition 

    Ambient Temperature:-30℃ ~ +55℃ 

    Relative Humidity:  ≤ 95% 

    Equipment Classification 

    Class Ii, Internal Power Supply, Type Cf 

    Ip Classification 

    Ipx4 

    Dimension 

    140mm (L) &Times;157mm(W) &Times;220mm(H) 

    Weight 

    1.8kg (Net Weight) 