SHORT SPECIFICATION
- 3 Drum
- 200 Liter
- Fully Automatic
- Chamber 316L
- 4mm Thickness
- Jacket 304
- 4mm Thickness
- = Optional - Double Door @ 60000 Extra
BENEFIT OF AUTOMATIC AUTOCLAVE MACHINE OVER MANUAL AUTOCLAVE MACHINE
In An Automated Autoclave, The Microprocessor Controls All The Processes Involved In Sterilization. It Is Also Provided With A Vacuum Pump.
VACUUM PUMP
A Water Ring Vacuum Pump Is Provided With All Our Automatic Autoclave For Following Reasons
1. Pre Vacuum: Before Exposure Of Steam, The Vacuum Pump (Available Only In Automatic System) Creates Vacuum Through A Pulsing Process (Filling Of Vacuum With Dry Steam), Thus Removing The Bio Burden From The Chamber And Increasing The Sterilization Potency Of The Process.
2. Post Vacuum: The Vacuum Pump Removes The Residual Moisture After Steam Exposure From The Chamber At The End Of Sterilization Cycle To Enhance The Dryness Of Resultant Load.
ADVANTAGES OF PLC AUTOMATION
A Programmable Logic Controller, PLC Or Programmable Controller Is A Digital Computer Used For Automation Of Electro- Mechanical Processes, Such As Control Of Machinery.
3. Pre-Programmed Cycles: Sterilization Of Varied Equipment’s Like Gloves, Linen, Gauze, Cotton, Instruments, Glassware Etc. Using Standard 121°c & 134°c Which Are Preprogrammed In The PLC/ Microprocessor The System Includes The Following Pre-Programmed Cycles; Additionally, Unlimited Number Of User Defined Cycles Can Also Be Programmed Into The Memory Of The System:
H.P.H.V. CYCLE
BOWIE & DICK TEST PROGRAM
TERMINAL CYCLE
LEAK TEST CYCLE
4. Control: During Sterilization, PLC Controls The Temperature, Moisture, Valve Control, And All Critical Time And Pressure-Based Parameters.
5. Temperature : Is Maintained Uniformly & Accurately (± 0.5°c). Continuous Display Of Temperature, Pressure, And Humidity On-Line Allows The User To Be Aware Of Cycle Status At All Times.
6. Reports: Due To The Advanced Software, Reports Are Saved In Data, Summary & Graph Formats For Every Batch Of Goods Sterilized.
7. Print: Users Can Reprint Reports As And When Required Using Any Printer. Digital Data Cuts Cost & Space, Ensures Safety Of Data, And Negates The Cost Of Expensive Thermal Printers.
8. HMI: The Operator Can Run A Machine Through HMI (Human Machine Interface). He Can Also Change The Programs According To The Hospitals Suitability.
9. Security: Advanced Multi-Level Password Protection Enhances Security. This Avoids Unauthorized Access.
20"x 38" Horizontal Fully Automatic Autoclave With vacuum system
Easy Return And Exchange Policy
Non Returnable
Installation
Free
Product Condition
New
Brand
Honmed
Warranty Policy :
One Year
Delivery Time :
15 To 25 Day's
TECHNICAL SPECIFICATION
Size :
500mm (Ø) X 950mm (L) Or (20” X 38”) (200 Liters/ 3 Drum Capacity)
Chamber:
SS 316 L (Thickness: -5 MM)
Jacket:
SS 304 L (Thickness: -4 MM)
Door:
SS 316 L (Thickness: -16 MM) Plates On Contact & Non-Contact Sides
Hinge:
SS
Central Locking Parts:
CI Casting, Wrinkle-Black Painted
Shooting Bolts
Tempered, Chrome Coated
Outer Insulation:
304 SS
Operating Pressure:
2.5kg/Cm2
Hydraulic Test Pressure:
4.0kg/Cm2
Operating Temperature
121 / 134°c / Bowie – Dick Cycle
Stand :
Ms Pipe
Operation:
Auto/Manual
Steam Generator:
SS 304 L (30 Liter)
Heating:
2 Industrial Immersion Heaters Of 6kw Each
Vacuum Pump:
1 H.P. Make: Kirloskar
Other Accessories:
Steam Trap, Pressure Switch, Float Switch, Temp Sensor, Pressure Transmitter, Pressure Gauge, Temp Gauge, Non-Return Valve, Rotex Solenoid Valves, Vacuum Pump Etc.
(1) Steam Trap: Forbes Marshall
(2) Pressure Switch: Danfos
(3) Pressure Transmitter: Honeywell
(4) Pressure Gauge: Baumer
(5) Temperature Gauge: Baumer
(6) Solenoid Valves: Recon/Jekon
Hardware:
Plc Make: Siemens
Hmi Make: Weintek, 7” Digital Display
Process Reports With Printer
Reports Are Generated By The Software For Each And Every Cycle In Detailed Report
SITE REQUIREMENT
Chamber Size
Carpet Size (Feet)
Water Drain
Water Inlet
Water Consumption
Vacuum Pump
Heater
3 Phase Wire Size
Load
Mcb
20” (Día) X 38” (Length)
3’(W)X4’(L)X 5’(H)
2-Inch, Gi Pipeline (Material)
½’ Inch Pipe With 2 Dividers
18 Liter Per Cycle & 40 Liter Boiler Capacity
1 Hp
6kw
(Qty:2)
4 Sq.Mm X 4 Core Copper Cable
15 Kw
63 Amp